Powering the Node
36
5.4
Connecting an External Power Supply
Apply only the input voltage range specified for the node in the
polarity indicated. Failure to do so could result in personal
injury and permanent damage to the node (
).
When the internal node switch is set for external power, the node may be directly powered by
the power supply specified for charging the node (or another regulated AC to DC power supply
with the appropriate output parameters,
see Operating Specifications on page 60
). It can also
be powered by an external battery or other regulated DC supply. The supply must deliver a
stable voltage between 3.2 and 9.0 V dc and be capable of sourcing at least 100 mA. External
battery holders and ruggedized outdoor housings that accommodate D cell batteries are
available for the G-Link
®
-LXRS
®
(
see Node Accessories on page 55
).
External power is applied through the power supply jack, Observe connection polarities, or the
node may be damaged.
Figure 27 - External Power Connections