9. Electromagnetic compatibility
Guidance and Manufacturer’s declaration-electromagnetic emissions
The Elexo is intended for use in the electromagnetic environment specified below . The customer or the user of Elexo should
assure that it is used in the following environment:
Emissions Test
Compliance
Electromagnetic Environment - guidance
RF emissions
CISPR11
Group 1
The Elexo uses RF energy only for its internal function . Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment .
RF emissions
CISPR11
Class B
The Elexo is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes .
Harmonic emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/
Flicker emissions
IEC 61000-3-3
Complies
Guidance and Manufacturer’s declaration-electromagnetic immunity
The Elexo is intended for use in the electromagnetic environment specified below . The customer or the user of the Elexo should
assure that it is used in the following environment:
Immunity Test
IEC 60601-1-2 Test
Level
Compliance Level
Electromagnetic Environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be made of wood, concrete, or ceramic
tile . If floors are covered with synthetic material, the
relative humidity should be at least 30% .
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality should be that of a typical
commercial or hospital environment .
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Mains power quality should be that of a typical
commercial or hospital environment .
Voltage dips, short
interruptions, and voltage
variations on power
supply input lines
IEC 61000-4-11
<5% Un
(>95% drop in Un)
for 0,5 cycle
40% Un
(60% drop in Un) for
5 cycles
70% Un
(30% drop in Un) for
25 cycles
<5% Un
(>95% drop in Un)
for 5 seconds
<5% Un
(>95% drop in Un)
for 0 .5 cycle
40% Un
(60% drop in Un) for
5 cycles
70% Un
(30% drop in Un) for
25 cycles
<5% Un
(>95% drop in Un)
for 5 seconds
Mains power quality should be that of a typical
commercial or hospital environment . Elexo is
equipped with a battery .
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment .
NOTE
Un is the a .c . mains voltage prior to application of the test level
Summary of Contents for 5100 5500
Page 2: ...Gebruikershandleiding Elexo 5100 5500 Douche toiletstoel...
Page 22: ...Instructions for use Elexo 5100 5500 Shower toilet chair...
Page 42: ...Mode d emploi Elexo 5100 5500 Chaise de douchetoilette...
Page 63: ...Elexo Chaise de douche toilette Mode d emploi 63 Fran ais Deutsch English Nederlands Notes...
Page 64: ...Benutzeranleitung Elexo 5100 5500 Dusch und Toilettenstuhl...
Page 86: ...86 Benutzeranleitung Elexo Dusch und Toilettenstuhl Notizen...