62
|
Mode d’emploi
| Elexo Chaise de douche-toilette
Recommended separation distance between
portable and mobile RF communications equipment and the Elexo
The Elexo is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled . The
customer or the user of the Elexo can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the Elexo as recommended below, according to the
maximum output power of the communications equipment .
Rated maximum output power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d =1,2 √P
80 MHz to 800 MHz
d =1,2 √P
800 MHz to 2,5 GHz
d =2,4 √P
0,01
0,12
0,12
0,24
0,1
0,38
0,38
0,76
1
1,2
1,2
2,4
10
3,8
3,8
7,6
100
12
12
24
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating
of the transmitter in watt (W) according to the transmitter manufacturer .
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies .
NOTE 2
These guidelines may not apply in all situations . Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people .
Summary of Contents for 5100 5500
Page 2: ...Gebruikershandleiding Elexo 5100 5500 Douche toiletstoel...
Page 22: ...Instructions for use Elexo 5100 5500 Shower toilet chair...
Page 42: ...Mode d emploi Elexo 5100 5500 Chaise de douchetoilette...
Page 63: ...Elexo Chaise de douche toilette Mode d emploi 63 Fran ais Deutsch English Nederlands Notes...
Page 64: ...Benutzeranleitung Elexo 5100 5500 Dusch und Toilettenstuhl...
Page 86: ...86 Benutzeranleitung Elexo Dusch und Toilettenstuhl Notizen...