Interpreting the Data
USRID-16W User Guide
24
l
A
º
symbol will be shown, if the USRID-16W was downloaded with LogTag
®
Analyzer.
l
A
•
symbol will be shown if an inspection mark was placed with the
START
/Mark
button.
l
A
•
symbol will be shown where the logger was plugged into a USB port at the time it
would otherwise have taken a reading.
USB Paused Readings
USB loggers cannot take a reading while plugged into a USB port. A gap is shown in the
graph where the USRID-16W was plugged in at the time it would otherwise have taken a
reading. The list shows
---
instead of the reading.
Legend
Shows the symbols for
download marks
,
inspections marks
and
paused marks
if they
appear in the readings.
Logger Statistics Overview
This section gives a brief overview of the temperature data collected during the trip. It
shows minimum and maximum values, when these occurred and also shows average,
standard deviation and MKT values.
Recorded Data Chart
The chart shows a graphical representation of the data during the trip. As part of the
USRID-16W configuration process you set the parameters that influence how the chart is
presented.
File Information
This section shows general information about the PDF file, such as generation time, date
and time formats used in the chart and the data list as well as the file name, which is
compiled from information about the data it contains:
LogTag_[serial_number]_[trip number]_[file creation date]_file creation time]_[OK or ALM].pdf
Other files that may be generated have the extensions *.csv and *.ltd.
Logger Information and Configuration
This section shows general information such as serial number, model number, trip
number, battery status and description. It shows how the logger was started, if a start
delay was active and the interval used for taking readings.
Temperature Alarms
This section summarizes the alarm trigger conditions and occurrences during the trip,
including:
www.