Operating the Logger
USRID-16W User Guide
20
Pressing
STOP
/Review
for a second time will show the
lowest recorded temperature during the trip, indicated
by
TRIP MIN
.
The example screen shows that the minimum
temperature recorded was -1.5 °C.
Pressing
STOP
/Review
again will show up to three more screens, one for each of the
configured alarm trigger conditions.
A screen for a specific alarm condition will only be shown if an alarm was triggered by this
threshold. If no alarm was triggered during the trip, only the MIN and MAX screens are
shown.
The additional alarm screens will show the alarm threshold values in descending order,
starting with the upper most alarm threshold.
In this example, all three alarms triggered, so all three screens are shown.
The first of the alarm screens shows the uppermost
alarm threshold, and the time that elapsed between the
start of the trip and the time the alarm was triggered.
In the example, the temperatures were equal/above +45
°C for one complete hour, and the alarm was triggered 17
days and 6 hours into the trip.
The second alarm screen shows the next threshold. In
the example, the temperatures were equal/above +30 °C
for a total of 10 hours, and the alarm was triggered 16
days and 21 hours into the trip.
The third alarm screen shows the lowest threshold. In
the example, the temperatures were equal/below -0.5 °C
for one complete hour, and the alarm was triggered 10
days and 4 hours into the trip.
Pressing
STOP
/Review
again starts the review cycle again
with the MAX screen.
Plugging the USRID-16W into a USB port
As soon as you plug the USRID-16W into a USB port, the word
USb
appears.
If you have configured the logger to generate files,
USb
remains turned on until all files
have been generated.
www.
.com
1.800.561.8187