Configuring the USRID-16W for
logging
USRID-16W User Guide
9
Configuring the USRID-16W for logging
Each USRID-16W model is delivered to you pre-configured with the prescribed WHO
recording profile, ready to be started.
The display is off. Briefly pressing the
STOP
/Review
button shows:
Although the logger is already configured, you can make a limited number of adjustments
to the configuration. These include:
l
Adding a description, which will show on the PDF report
l
Changing the appearance of the PDF report
l
Changing the language in which the PDF report will be printed
l
Selecting which files are generated at the end of the trip
This is done using LogTag
®
Analyzer software, which can also be used for downloading and
analyzing data, using the following steps:
l
Start the LogTag
®
Analyzer software.
l
Remove the protective cap from the logger's USB connector, plug it into a USB
socket on your computer and wait for the drivers to be installed
3
.
l
From the menu click
LogTag - Configure
; LogTag
®
Analyzer will display the
configuration options for connected loggers.
Standard Configuration Options
The only parameter in the standard configuration options you are permitted to customize
is the
Description
.
All other parameters are preset to the requirements of the WHO for shipping indicators:
Start Option:
Push Button
Number of days to record:
20
Recording Interval:
5 minutes
Start delay:
1 hour
3
You can configure more than one USRID-16W at the same time, however it is practical to limit the number of units to about 6,
using a powered USB hub.