Standard Configuration Options
TREL30-16 User Guide
14
previous day is shown (
) to indicate the alarm was registered against what is now the
previous day. When midnight passes next, this marker will move to and so on.
A note on alarm re-triggering:
As soon as an alarm is triggered, the corresponding
delay time (but not any others) resets to zero and alarm processing starts again.
The alarm processing for all other alarm delays is not affected. Therefore,
accumulative or consecutive alarms will re-trigger, if the alarm conditions are met
again, and the Alarm Indicator ( ) and the Day Alarm Marker (
) will be shown,
even if any previous alarm was cleared.
Clearing an Alarm does not reset any of
Advanced Configuration Settings
for additional configuration settings. These settings decide how
some of the elements are displayed on the unit's own display and set certain options
specific to the TREL30-16.
TREL30-16 advanced configuration screen in LogTag
Parameters that influence the appearance of the display are:
Pause alarm/statistics processing (see
)
l
Temperature display unit (see
)
l
Switch off display after 30 seconds (Power save, see
)
l
showing the number of days on the display, for which statistics were collected (see
Total Number of Days Statistics Collected on page 28
)
l
allowing a user to reset the trip's minimum and maximum values on the display
during recording
(see Resetting the Trip Minimum/Maximum Temperatures on
Parameters influencing specific behavior of the TREL30-16 are:
l
clearing and resetting alarms when the
START
/CLEAR/STOP
button is pressed