13.
Press the
►
to “Configure”, by default
the screen should read as follows.
14.
The “Configure Gages” screen needs to be set as follows. Failure to set this screen properly
may cause undesired results. The top screen shows a standard LMI 241 configuration, the
bottom screen shows a standard 241-B configuration.
Configure Gage setup for LMI 241 gap
“Gage Port G3” must
be matched to “Master
with G2” to calibrate
both flush and gap at
the same time.
This configuration will
produce a negative value
when the LMI 241 is
extended from the master
point (nominal). To make
the LMI 241 read positive
when extended from the
nominal change the scale
to “-10”
Configure Gage setup for LMI 241-B gap
This configuration will
produce a 3mm value
when the LMI 241 is at
the master point
(nominal).
“Gage Port G3” must
be matched to “Master
with G2” to calibrate
both flush and gap at
the same time.
Form: CA 030
December 28,
2004
R:\Quality\Calibration Instructions\CA 030.doc
Page 7 of