The “Configure Gages” screen needs to be set as follows. Failure to set this screen properly
may cause undesired results.
8.
If changes to this screen are needed, press
▲
or
▼
to highlight either “Type”, “Master Type”,
“Master With”, or “Zero Before Read” then press <enter> to toggle the different choices.
9.
To make changes to the “Scale” or “Zero Master” press
▲
or
▼
to highlight “Scale” or “Zero
Master” and press <enter>. This will bring up the numeric keypad. Key in the new value and
press
►►
to accept.
“Gage Port G2” must
be matched to “Master
with G3” to calibrate
both flush and gap at
the same time.
If the “Gage Port” is
set to “G2a” then the
“Master” needs to be
G3a
Form: CA 030
December 28,
2004
R:\Quality\Calibration Instructions\CA 030.doc