Gap Setup 

This configuration will produce a negative value for flush when the tip extends beyond the 

nominal (master) point. To reverse the values enter -10 in step 10. 

 
 
11.

 

To assign the gage file name, press 

 or 

 to highlight “G3”* in the “Gage List” and press 

<enter> on the collector. 

 

*G3 has the ability to support “sub files”. These are files that allow multiple styles of gages to be 
used by the same collector. To create sub files highlight G2 and press 

►►

, this will create a sub 

file of G3a. Each press of 

►►

 creates another sub file. The balance of this instruction will be 

based on G3 

 
12.

 

Use the 

,

,

, or 

 to highlight the first character of the new file name then press 

<enter>, repeat the process until the gage file name is spelled. Then press 

►►

 to accept the 

new name. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Form:  CA 030 

December 28, 

2004 

R:\Quality\Calibration Instructions\CA 030.doc 

Page 6 of  

13 

 

Summary of Contents for 241 Series

Page 1: ...01 2 3 4 1 Required Equipment 1 LMI 440 or ASIDataMyte 501 2 LMI 241 or 241 B transducer 3 LMI 720 master block 4 LMI 6025 6 pin 6 pin cable This instruction will outline I GAGE CONFIGURATION 2 Flush...

Page 2: ...ended to store a copy of the gage files onto a personal computer or laptop Consult the collector manual or if purchased the TranSend manual for further details Flush Setup This configuration will prod...

Page 3: ...that allow multiple styles or configurations of gages to be used by the same collector To create sub files highlight G2 and press this will create a sub file of G2a Each press of creates another sub f...

Page 4: ...th or Zero Before Read then press enter to toggle the different choices 9 To make changes to the Scale or Zero Master press or to highlight Scale or Zero Master and press enter This will bring up the...

Page 5: ...accept the new value 11 Once the Configure Gages screen is set press and a pop up will appear if any changes were made Highlight Save to current gage and press enter if the changes were intentional If...

Page 6: ...r G3 has the ability to support sub files These are files that allow multiple styles of gages to be used by the same collector To create sub files highlight G2 and press this will create a sub file of...

Page 7: ...I 241 gap Gage Port G3 must be matched to Master with G2 to calibrate both flush and gap at the same time This configuration will produce a negative value when the LMI 241 is extended from the master...

Page 8: ...cale or Zero Master and press enter This will bring up the numeric keypad Key in the new value and press to accept 17 Key in the desired value buy using the or to highlight the first number press ente...

Page 9: ...e in Configure Gages highlight Cancel and press enter Configuration of the LMI 440 or ASIDataMyte 501 is complete II MASTERING INSTRUCTIONS 1 Connect the transducer to Gage Port 2 3 of the data collec...

Page 10: ...n the collector 6 Master will be highlighted in screen header and G2 is identified as Gage Port If G2 is not the Gage Port press the or until G2 appears Master highlighted G2 is the Gage Port 7 Extend...

Page 11: ...ap slot With Master Zero highlighted on the collector press enter Mastering for the 241 BW series is complete III Verification of the Mastering This process can be performed to verify the mastering of...

Page 12: ...he result must be 10 00mm 0 04mm This example step 1 is 5 002 and step 2 is 5 002 added together without the minus sign equals 10 004mm 4 Insert the gage into the master block and observe the value in...

Page 13: ...erve the value in the Master Gages screen of Gage Port G3 in the collector LMI 241 reading LMI 241 B reading Value reading for Gap retracted 8 Add the results of step 6 and step 7 together disregard t...