TROUBLESHOOTING TIPS
PROBLEM
LIKELY CAUSE
METHOD OF CORRECTION
After pressing the
Button no indication
on the display.
Discharge of batteries.
Polarity of batteries is not
observed.
Battery terminals are
contaminated
Power source not
plugged in an electrical
outlet.
Replace all batteries for
new ones.
Install batteries correctly.
Clean the terminals with
dry cloth.
Plug the power source into
an electrical outlet.
Infl ation is stopped
and resumed.
Automatic re-infl ation
to obtain correct
measurements.
Perhaps you talk or move
your arm during the
measurement?
See MEASUREMENT
PROCEDURE
Calm down and repeat
the measurement.
Every time the blood
pressure is diff erent.
Measurements are
too low/high.
Check that the arm cuff is
level with your heart?
Check that the arm cuff is
applied correctly?
Perhaps your arm muscles
are tough?
Perhaps you talk or move
your arm during the
measurement?
Take the correct position
for measurement.
Take the correct position
for measurement.
Calm down, apply the
arm cuff correctly.
Keep silence and quiet
during measurement.
Measurements of
the pulse rate are
too high/low.
Perhaps you talk or move
your arm during the
measurement?
Perhaps you make
measurement directly after
physical load?
Keep silence and quiet
during measurement.
Repeat the measurement
at least in 5 min.
Impossibility
to make a large
of number of
measurements.
Application of poor
batteries.
Use only alkali batteries
of well-known
manufacturers.
If regardless of the above recommendation you are unable to get correct
measurement results, stop the use of this Device and apply to a maintenance
organization (addresses and telephones of authorized organizations may
be found in the warranty card). Do not try to adjust the internal mechanism
by yourself.