LY71 Initial Setup Manual

1/16

LY71 

Initial Setup Manual

Contents

Page

Preparations before making the initial settings (checking conditions of use)

2 to 4

How to set up 

Basic Settings

5 to 7

How to set up 

Advanced Settings 

(continued from Basic Settings)

8 to 10

How to set up the BCD unit (optional)

11

How to set up a comparator unit (optional)

12

Factory Default (All Clear)

13

Appendix 1: Front panel, Alarm Indication

14

Appendix 2: Adapter connections (scales and gauges)

15 to 16

Initial Setup Flow

Basic 

Settings

Advanced 

Settings

Table of contents

STEP

1~3

STEP

4~5

STEP

4~5

STEP 

1~3

START

Finished

Ver.2 (2020.11) CS&S

LY71

Advance 

Preparation

Check the conditions of use

Option

STEP

6~7

Note: 

Not required if the option is not used

(Note)

STEP

6~7

STEP

6~7

LZ71-B

LZ71-KR

Confirmation: software version 1.15

Summary of Contents for LY71

Page 1: ...optional 11 How to set up a comparator unit optional 12 Factory Default All Clear 13 Appendix 1 Front panel Alarm Indication 14 Appendix 2 Adapter connections scales and gauges 15 to 16 Initial Setup...

Page 2: ...selections increase the number of options Length 0 05 to 100 m See Table 1 2 and 3 or Angle 1 second to 1 degree Angular resolution 1sec to 1 degree when using a rotary scale Basic Setting Items Note...

Page 3: ...Latch function PAUSE Pause function General purpose input One external signal input is assigned to display A B and C axes Several functions can be operated by external signals but only one function ca...

Page 4: ...FF 0 L ON Negative logic 1 L ON 0 H OFF The logic of the sign can be set Positive logic H OFF L ON Negative logic L ON H OFF The READY signal is fixed in negative logic Enable L ON Disable H OFF LGC L...

Page 5: ...s A Addition of 2 axes Data selection C Current value R Maximum value I Minimum value P P P value Display switching The method of displaying the comparator value can be selected Manual You can check t...

Page 6: ...first and second axes Independent display axis added display etc How to set up Basic Settings 1 3 Step 1 COUNTRY Destination When you want to change the basic settings The following display appears No...

Page 7: ...t axis How to view extended settings For extended settings press SART key when the input resolution of each axis is displayed 2 m 100 m 50 m 25 m 20 m 0 05 m 1 degree How to set up Basic Settings 2 3...

Page 8: ...will be After the second time the screen is displayed when the power is turned on Press key The display returns to normal This completes the basic settings Step 3 Setup key Enter key Select axis key...

Page 9: ...the advanced settings If you have already completed the basic settings and want to reconfigure only the advanced settings you can do the same If you select the count display you cannot make basic set...

Page 10: ...n can be set to prevent accidental operation It is possible to save the display at power off and display the value at power on Flickering of the smallest digits can be suppressed in two stages If no o...

Page 11: ...lay Time ms No delay Output data state when an alarm occurs and when the power is turned on OFF Output BCD output mode can be set B axis You can set the logic of the output signal The logic of BCD dat...

Page 12: ...election keys for each axis to set the functions A axis Set the output mode of the data Outputs the displayed value Latch the comparator output 0 5s when passing the positional point Specify what to j...

Page 13: ...her than the explanation 1 Hold down key and key at the same time to start the power supply The model name of the counter is displayed Factory setting All clear Reference key Connect the AC adapter to...

Page 14: ...r Overflow Power failure Note When using an adapter connection SZ no speed override indication is shown but rather an error message Light Blinking Front panel No Name No Name No Name ABS lamp REF key...

Page 15: ...crews Use screws to secure it in place Use screws to secure it in place SZ70 1 Use screws to secure it in place Scale Head Resolution Adapter Counter SR128 GB A 0 5 m SZ70 1 LG20 LH70 71 71A 72 LY71 7...

Page 16: ...screws to secure it in place SZ70 1 Counter unit Screws DK series CE29 Use screws to secure it in place Gauge Resolution Adapter 1 Adapter 2 Counter DG B 0 5 m DZ51 SZ70 1 LG20 LH70 71 71A 72 LY71 72...

Reviews: