Set up the Counter
248083395-1 Rev 9
4-23
WARNING:
Step 18 is
very important to prevent
network address errors
on the PC when the
process is complete.
18.
Clear the ARP Cache by typing
arp –d *
and press
Enter
.
If this is
not done before quitting this process, the PC may create “ghost
IPs” on the LAN and cause serious problems with the network.
19.
Type
Exit
to quit the command console.
Connect the
SOLAIR
1100LD to
Production
LAN
Connect Ethernet Cable to Instrument
It is presumed the previous steps were performed on the SOLAIR in an
Off-LAN state (not connected to a production network). Power the
SOLAIR OFF and disconnect any previously attached cabling.
Attach the network cable that will be used for production to the
SOLAIR and the other end to the production network receptacle.
Apply power to the instrument and check the port LEDs to verify the
connection is working.
When all steps are completed, the SOLAIR will supply data to the LAN
connection. After approximately 60 seconds, monitoring equipment
and software should “see” the instrument and be able to retrieve data
from it.
For additional information on data provided by the 1100LD, see “Data
View Buffer Screen” in “Operating Counter” on page 6-53 and the
monitoring software’s Operating Manual for detail information
regarding downloading of this data.
Contact Lighthouse Worldwide Solutions Technical Support at 1-800-
945-5905 (Toll Free USA) or 1-541-770-5905 (Outside of USA) for
additional information or further assistance.
Summary of Contents for SOLAIR 1100LD
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions SOLAIR 1100LD Airborne Particle Counter Gen E Operating Manual...
Page 6: ......
Page 12: ...Lighthouse SOLAIR 1100LD Operating Manual t vi 248083395 1 Rev 9...
Page 14: ...Lighthouse SOLAIR 1100LD Operating Manual ii 248083395 1 Rev 9...
Page 20: ...Lighthouse SOLAIR 1100LD Operating Manual 2 4 248083395 1 Rev 9...
Page 58: ...Lighthouse SOLAIR 1100LD Operating Manual 4 24 248083395 1 Rev 9...
Page 64: ...Lighthouse SOLAIR 1100LD Operating Manual 5 6 248083395 1 Rev 9...
Page 146: ...Operate SOLAIR 248083395 1 Rev 9 6 82...
Page 158: ...Lighthouse SOLAIR 1100LD Operating Manual A 2 248083395 1 Rev 9...
Page 180: ...Lighthouse SOLAIR 1100LD Operating Manual D 2 248083395 1 Rev 9...
Page 187: ...00...
Page 188: ......