Lighthouse SOLAIR 1100LD Operating Manual
6-78
248083395-1 Rev 9
Sample Printouts of Standard Reports
Figure 6-102, Figure 6-103 and Figure 6-104 are examples of a Federal
Standard Ft3, ISO 14644-1 and EU GMP 2009 reports.
Note:
If a recipe is
loaded, the name of the
recipe will be printed on
the report.
Figure 6-102 Sample Federal Standard ft
3
report
*************************
Federal Standard Ft^3
**SOLAIR 1100LD**
Serial #: 130705002
Targeted Class: 10000
Room Area: 10ft^2
Room Stat: Operational
Air Flow: Unidirect
Min Loc: 2
Min Samples/Room: 5
Recipe: FED_TEST
-----
07/10/2013, 14:35:47
-----
Particle Size: 0.10
Cumulative, p/ft^3
Vol Req: 0.100 ft^3
Concen Limit: 30800
-----
Loc Samples AvgConcen
LOC001 5 8003.5
LOC002 5 6524.5
-----
Mean: 7264.0
StdDev: 1045.8
StdError: 739.5
95% UCL: 11930.2
Min Vol: 1.467 ft^3
-----
Particle Size: 0.15
Cumulative, p/ft^3
Vol Req: 0.100 ft^3
Concen Limit: 10000
-----
Loc Samples AvgConcen
LOC001 5 328.4
Summary of Contents for SOLAIR 1100LD
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions SOLAIR 1100LD Airborne Particle Counter Gen E Operating Manual...
Page 6: ......
Page 12: ...Lighthouse SOLAIR 1100LD Operating Manual t vi 248083395 1 Rev 9...
Page 14: ...Lighthouse SOLAIR 1100LD Operating Manual ii 248083395 1 Rev 9...
Page 20: ...Lighthouse SOLAIR 1100LD Operating Manual 2 4 248083395 1 Rev 9...
Page 58: ...Lighthouse SOLAIR 1100LD Operating Manual 4 24 248083395 1 Rev 9...
Page 64: ...Lighthouse SOLAIR 1100LD Operating Manual 5 6 248083395 1 Rev 9...
Page 146: ...Operate SOLAIR 248083395 1 Rev 9 6 82...
Page 158: ...Lighthouse SOLAIR 1100LD Operating Manual A 2 248083395 1 Rev 9...
Page 180: ...Lighthouse SOLAIR 1100LD Operating Manual D 2 248083395 1 Rev 9...
Page 187: ...00...
Page 188: ......