Programming
248083310-1 Rev 3
4-11
secs=2: ch1-2: 1115, 382 , bk= 429, alarm alert.
secs=3: ch1-2: 1595, 502 , bk= 429, alarm alert.
secs=4: ch1-2: 2029, 637 , bk= 429, alarm alert.
secs=5: ch1-2: 2381, 738 , bk= 429, alarm alert.
Sample: ch1-2: 2381, 738 Range1= 10000, Range2= 5000.
Alarms: Threshold= 1000, Channel= 1, # in a row=2
SET ALARM SUPPRESSION
<ae#>
Sets the alarm suppression for the configured alarm channel. The
number of alarm conditions have to be greater than # in order to turn on
the Alarm Relay.
For example, if Alarm Suppress is set to 2, then the device won’t report
an alarm condition via the external alarm output or the flashing service
light until the 3rd consecutive alarm condition.
Type:
<ae2>
Response:
<ae2>Suppress Alarms = 2
Alarm Alert with Alarm Suppression Example:
In this example, the Alarm Suppress is set to 2.
<ae2>Suppress Alarms = 2
secs=1: ch1-2: 0, 0 , bk= 429
secs=2: ch1-2: 681, 430 , bk= 2558
secs=3: ch1-2: 1629, 886 , bk= 444
Sample: ch1-2: 1629, 886 Range1= 10000, Range2= 5000.
Note:
The first line
starting with "Alarms" (in
boldface) shows what the
current alarm threshold is,
which channel is set for
alarming and how many
alarm conditions in a row
have occurred.
Alarms: Threshold= 1000, Channel= 1, # in a row=1
secs=1: ch1-2: 828, 517 , bk= 1640
secs=2: ch1-2: 1861, 1230 , bk= 434
secs=3: ch1-2: 3149, 2177 , bk= 449
Sample: ch1-2: 3149, 2177 Range1= 10000, Range2= 5000.
Alarms: Threshold= 1000, Channel= 1, # in a row=2
secs=1: ch1-2: 450, 126 , bk= 434
secs=2: ch1-2: 1279, 530 , bk= 444
secs=3: ch1-2: 1875, 678 , bk= 439
Summary of Contents for REMOTE 3012PN
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE 2 PN Series Airborne Particle Counter Operating Manual...
Page 6: ......
Page 10: ...Lighthouse REMOTE 2 PN Series Operating Manual t iv 248083310 1 Rev 3...
Page 12: ...Lighthouse REMOTE 2 PN Series Operating Manual ii 248083310 1 Rev 3...
Page 14: ...Lighthouse REMOTE 2 PN Series Operating Manual 1 2 248083310 1 Rev 3...
Page 42: ...Lighthouse REMOTE 2 PN Series Operating Manual 4 14 248083310 1 Rev 3...
Page 46: ...Lighthouse REMOTE 2 PN Series Operating Manual 5 4 248083310 1 Rev 3...
Page 50: ...Lighthouse REMOTE 2 PN Series Operating Manual A 2 248083310 1 Rev 3...
Page 52: ...Lighthouse REMOTE 2 PN Series Operating Manual LOF 2 248083310 1 Rev 3...
Page 56: ...Lighthouse REMOTE 2 PN Series Operating Manual I 4 248083310 1 Rev 3...
Page 57: ...00...
Page 58: ......