Programming
248083310-1 Rev 3
4-9
<gv> Version: 010
SET CHANNEL RANGES
Note:
Range 1 should
always be greater than or
equal to Range 2; Range 2
should never be greater
than Range 1.
<ra#>
-- sets upper range of channel 1 where # is the range
value
<rb#>
-- sets upper range of channel 2 where # is the range
value
Sets upper limit to counting range. The upper limit will be equal to
20mA, and the lower range will always be 0 counts = 4 mA. If the
counts exceed range limit, the counts will be set to the range limit. The
maximum range is 10,000,000.
Type:
<ra10000>
Response:
<ra10000>Range1= 10000.
Type:
<rb7500>
Response:
<rb7500>Range2= 7500.
SET SAMPLE TIME (seconds)
Note:
Sample times <5
seconds are not
recommended.
<ta#>
Sets sample time in # seconds, where # >1 and nnnn < 3600.
Type:
<ta30>
Response:
<ta30>Sample Time = 30 secs
SET ALARM CHANNEL
<aa#>
Summary of Contents for REMOTE 3012PN
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE 2 PN Series Airborne Particle Counter Operating Manual...
Page 6: ......
Page 10: ...Lighthouse REMOTE 2 PN Series Operating Manual t iv 248083310 1 Rev 3...
Page 12: ...Lighthouse REMOTE 2 PN Series Operating Manual ii 248083310 1 Rev 3...
Page 14: ...Lighthouse REMOTE 2 PN Series Operating Manual 1 2 248083310 1 Rev 3...
Page 42: ...Lighthouse REMOTE 2 PN Series Operating Manual 4 14 248083310 1 Rev 3...
Page 46: ...Lighthouse REMOTE 2 PN Series Operating Manual 5 4 248083310 1 Rev 3...
Page 50: ...Lighthouse REMOTE 2 PN Series Operating Manual A 2 248083310 1 Rev 3...
Page 52: ...Lighthouse REMOTE 2 PN Series Operating Manual LOF 2 248083310 1 Rev 3...
Page 56: ...Lighthouse REMOTE 2 PN Series Operating Manual I 4 248083310 1 Rev 3...
Page 57: ...00...
Page 58: ......