Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual
5-40
5HY
Note:
When the buffer
has wrapped, the record
that is first displayed in the
Data screen is the first
record in the data buffer.
This may not be Rec#1.
If you press the single
down arrow button once,
the instrument will display
the last record in the data
buffer.
•
Rec#
- Identifies which record is currently viewed.
•
*Recs
- Displays how many records are currently stored in the
HANDHELD’s buffer. When there is an asterisk (*) next to the
word “Recs”, this means that the 3000 record data buffer has
wrapped. The basic concept is First In, First Out (FIFO).
•
Size
- Lists the channel sizes configured on the instrument.
•
PM
- Indicates Particulate Matter, expressed in micrograms per
cubic meter (ugr/m3). PM is calculated by summing the mass
concentration levels for all channels smaller than the indicated
channel. For example, the PM 1.0 = (Mass Concentration of 0.3) +
(Mass Concentration of 0.5). PM 10.0 = the sum of the Mass
Concentration levels of the 0.3 + 0.5 + 1.0 + 2.5 + 5.0
m channels.
Note that the Mass Concentration of the 10.0
m channel is not
included. It is, however, included in the TPM value, see TPM
below.
Figure 5-48 can be used to aid understanding of PM. Note that the
1.0 value, 6.41, is the sum of the values in the ugr/m3 entries for the
0.3 and 0.5
m channels (2.12 + 4.29 = 6.41).
•
TPM
- Total Particulate Matter, expressed in micrograms per cubic
meter, is the sum of mass concentration levels of all channels.
•
ugr/m3
- Lists the Mass Concentrations for each individual
channel, expressed in micrograms per cubic meter (ugr/m3).
•
Density
- The mathematical formula used to compute Mass
Concentration (simplifies to Mass = Density * Volume) assumes a
density value of Carbon, which is 2.5 grams per milliliter. The
Density factor may be changed, if desired, using the LMS XChange
program.
•
Loc
- Indicates the location at which the data record was recorded.
The Location name listed is the alphanumeric label that was saved.
•
Smpl
- Indicates the Sample Time (HH:MM:SS) at which the data
record was sampled.
•
Date
- Indicates the instrument date on which the data was
recorded. The date will display in the format selected in the
configuration (MM/DD/YYYY, DD/MM/YYYY or YYYY/MM/
DD).
•
Time
- Indicates the instrument time at which the data was
recorded (HH:MM:SS).
Summary of Contents for Handheld 3016 IAQ
Page 1: ...OPERATING MANUAL Handheld3016IAQ P A R T I C L E C O U N T E R...
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions HANDHELD 3016 IAQ Particle Counter Gen E Operating Manual...
Page 6: ......
Page 12: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual ii 5HY...
Page 14: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 1 2 5HY...
Page 18: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 2 4 5HY...
Page 78: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 4 50 5HY...
Page 128: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual A 2 5HY...
Page 130: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual B 2 5HY...
Page 135: ...00...