LifePort Kidney Transporter Operator’s Manual 755-00002 Rev R
37
CAUTION:
Use only grounded electrical connections. Connect LifePort Kidney Transporter to a
grounded electrical outlet rated for voltage and amperage according to the labeled ratings on the product
back panel. If there is any question about the ground integrity, operate LifePort Kidney Transporter from
internal power.
CAUTION:
You may remove AC mains power by unplugging the Power Cord from the back of the unit.
Exercise care when choosing the location of your LifePort Kidney Transporter so that removal of Power
Cord is not difficult.
CAUTION:
Do not allow cleaning solutions to enter the rear panel electrical connectors, the ventilation
holes, or the battery area.
CAUTION:
Use only cables and accessories approved by Organ Recovery Systems. Non-approved
cables and accessories may damage the system or interfere with accuracy. For information, contact the
Organ Recovery Systems 24/7 Perfusion Helpline.
CAUTION:
Do not substitute the Power Cord. Use only the Power Cord supplied from Organ Recovery
Systems. For information, contact the Organ Recovery Systems 24/7 Perfusion Helpline.
CAUTION:
Do not substitute batteries. Use only LifePort Kidney Transporter batteries from Organ
Recovery Systems. For information, contact the Organ Recovery Systems 24/7 Perfusion Helpline.
CAUTION:
Extended storage may damage the batteries.
CAUTION:
Replace the batteries only one at a time to ensure that LifePort Kidney Transporter will
continue to operate.
CAUTION:
Use precautions when lifting. A fully loaded LifePort Kidney Transporter weighs 45 lbs (20.4
kg). Use proper lifting practices to avoid injury.
CAUTION:
Do not immerse LifePort Kidney Transporter.
CAUTION:
Ice Container and Ice Container Lid are reusable parts of the LifePort Kidney Transporter.
Do not dispose of them.