LifePort Kidney Transporter Operator’s Manual 755-00002 Rev R
7
Access the batteries through the Battery Door on the LifePort Kidney Transporter External Connections Panel.
Each battery is designed to slide in and out of the slots provided. When inserted properly, the battery should be
flush with the slot panel, with the pull-tab visible and available for removing the battery. If the battery does not
push flush, it may be in the wrong orientation. Turn the battery 180 degrees and try again.
The following tips will help you obtain maximum life and serviceability from the batteries.
• Always replace the Battery Door. LifePort Kidney Transporter must not be operated or shipped without
the Battery Door in place.
• LifePort Kidney Transporter will recharge the batteries whenever it is plugged in. Plug in LifePort
Kidney Transporter whenever not in transit to keep the batteries at the highest possible charge. It takes
approximately five hours to completely
recharge all four batteries.
NOTE:
Keep extra charged batteries on hand when long transport times are anticipated or when
successive uses of LifePort Kidney Transporter with short turnaround times are expected.
• During storage of LifePort Kidney Transporter without connection to AC mains, the batteries will slowly
drain. After 30 days without charging, the batteries could have little or no charge and will need a full
five-hour
recharge.
• For periods of storage for longer than 30 days, remove the batteries from LifePort Kidney Transporter.
CAUTION:
Extended storage may damage the batteries.
Safe Disposal of LifePort Kidney Transporter and LifePort Batteries
Lithium-ion batteries must be disposed of according to local regulations. For safe disposal of LifePort Kidney
Transporter or LifePort Kidney Transporter batteries, call the Organ Recovery Systems 24/7 Perfusion Helpline
to arrange for pickup from your facility.
LifePort Kidney Transporter Disposable Products
Single-use Disposables, an integral part of the LifePort Kidney Transporter system, are used to maintain the
kidney and perfusate under aseptic conditions, to connect the kidney to the Perfusion Circuit, and to help
maintain aseptic conditions while working inside the Perfusion Circuit. Each LifePort Kidney Transporter
Disposable is factory sterilized and delivered in a sterile pack.
WARNING:
For single use only. Do not reuse, reprocess, or resterilize. Reusing, reprocessing,
or resterilization of single-use devices creates a potential risk of patient or user infections due to
contamination. This contamination may lead to injury, illness, or other serious patient complications.
Discard any unused portion of the product.
NOTE:
To reorder LifePort Kidney Transporter Disposables, please contact Organ Recovery Systems.
LifePort Kidney Transporter Disposable Cannulas
LifePort Kidney Transporter Disposable cannulas attach LifePort Kidney Transporter Disposable Perfusion
Circuit to the kidney’s renal artery. A large range of cannula types and sizes are available, making it possible to
choose the cannula most compatible to the anatomy of the kidney.
LifePort Kidney Transporter Disposable Sterile Drape
The LifePort Kidney Transporter Disposable Sterile Drape is used to aid in maintaining aseptic conditions while
working within the LifePort Kidney Transporter Disposable Perfusion Circuit.