LifePort Kidney Transporter Operator’s Manual 755-00002 Rev R
25
CAUTION:
Accessory equipment connected to the Data Port must be certified IEC 62368 for data
processing equipment. Furthermore, all configurations shall comply with the systems standard in cl. 16
of IEC60601-1. Any person who connects additional equipment to the Data Port configures a
medical system and is therefore responsible for ensuring that the system complies with the system
standard in cl. 16 of IEC60601-1. If in doubt, consult the Organ Recovery Systems 24/7 Perfusion
Helpline.
Using A Flash Drive
If a computer is not readily available for evaluating the data file, you can download the file onto a flash drive
and evaluate the data on a computer when one is available.
1. Power on LifePort Kidney Transporter.
NOTE:
If LifePort Kidney Transporter does not have a Perfusion Circuit installed, press
STOP
to clear
the “Sensor Not Connected” error and press
OK
.
2.
Insert USB flash drive into USB-A port on LifePort Kidney Transporter.
3. Use the arrow buttons to select
DOWNLOAD FILE
.
4.
Use arrow buttons to select file to download.
5. Press
OK
and
SAVE
. Top display will flash
SAVING FILE
until complete. When download is complete,
display will return to download file screen.
6.
You can download additional files, if desired, or use the arrow buttons to select
DONE
and press
OK
.
7.
Remove the USB flash drive.