755-00002 Rev R
LifePort Kidney Transporter Operator’s Manual
26
Troubleshooting and Diagnostics
Most problems that you encounter in operating LifePort Kidney Transporter will be easily solved. The first thing
to check when troubleshooting is to make sure that power is available from either the batteries or through
the Power Cord plugged into a standard electrical outlet. If the power light comes on but LifePort Kidney
Transporter still does not work, check the following guide.
Troubleshooting Procedures
Trouble
Probable Cause
Action
No Power
Dead Batteries
Outlet
Tripped Circuit
Breaker
1. Replace with fresh batteries or plug into an external power
supply. Make sure batteries are fully charged before using.
2. Make sure the outlet has power.
3. Reset breaker by pressing the button on the External
Connections Panel located on the back of LifePort Kidney
Transporter.
If problem is not resolved, please call Organ Recovery Systems
24/7 Perfusion Helpline.
Beeping or
Flashing Display
Beeping or Flashing
Display, Accompanied
by an Error Message
.
If problem is not resolved, please call Organ Recovery Systems
24/7 Perfusion Helpline.
Missing/Incorrect
Display
Display or Internal
Computer Error
1. Power
OFF
.
2. Power
ON
.
If problem is not resolved, call Organ Recovery Systems 24/7
Perfusion Helpline.
Leaking
Perfusate
Perfusion Circuit Lids
Not Tightened
Defective Perfusion
Circuit
1. Relatch the Perfusion Circuit Lids and look for any leaks
near the gaskets.
2. Replace the Perfusion Circuit. Call Organ Recovery
Systems 24/7 Perfusion Helpline to return Perfusion Circuit
for investigation.
If problem is not resolved, call Organ Recovery Systems 24/7
Helpline.
Leaking Coolant
Broken Ice Container
or Seal
Observe the Ice Container for any damage.
If damaged or if problem is not resolved, call Organ Recovery
Systems 24/7 Perfusion Helpline.
Unresponsive
Buttons
Internal Lock-up
1. Power
OFF
.
2. Disconnect Power Cord.
3. Remove all batteries.
4. Wait 30 seconds.
5. Return batteries to LifePort Kidney Transporter
6. Power
ON
.
If problem is not resolved, call Organ Recovery Systems 24/7
Perfusion Helpline.
Blank Display
Display or Internal
Computer Error
1. Power
OFF
.
2. Power
ON
.
If problem is not resolved, call Organ Recovery Systems 24/7
Perfusion Helpline.