LifePort Kidney Transporter Operator’s Manual 755-00002 Rev R
29
Error Message
Probable Cause
Actions
Too Cold
Incorrect Coolant
Environment Conditions
Too Cold (Ice Container
temperature has dropped
below 0.5°C)
1.
Check that ice and water only are used to fill
Ice
Container.
2. Move LifePort into a warmer environment.
If problem is not resolved, call Organ Recovery Systems
24/7 Perfusion Helpline.
Too Much
Pressure
High G Forces Created
During Transit
Cushion or reduce impact.
If problem is not resolved, call Organ Recovery Systems
24/7 Perfusion Helpline.
Too Warm Add
Ice
Ice Container Temperature
Above 8°C
1. Replenish ice as soon as possible.
2. Allow time for the temperature to read below 8°C
and press
INFUSE
to restart perfusion.
If problem is not resolved, call Organ Recovery Systems
24/7 Perfusion Helpline.
Upstream
Bubbles
Air Persisting in Upstream
Bubble Detector
Check Perfusion Circuit for leaks.
If problem is not resolved, call Organ Recovery Systems
24/7 Perfusion Helpline.
Watchdog
Internal Error
Call the Organ Recovery Systems 24/7
Perfusion Helpline.
Power On Self Test (POST)
On each power up, LifePort Kidney Transporter performs a Power On Self Test or “POST”. The CPU within
LifePort Kidney Transporter checks its memory functions, temperature sensors, Bubble Detectors, and its
internal failure routines. In the unlikely event that one of these tests fail, LifePort Kidney Transporter will display
“POST failure” and it will list the POST error message as shown in the table below. Should one of these errors
occur, remove all power to LifePort Kidney Transporter by reinstalling the batteries and Power Cord. If the
POST message continues to display, call the Organ Recovery Systems 24/7 Perfusion Helpline.
POST Error Message
Indication
POST AT error
Ice Bucket temperature error
POST failure
Power on self test failed
POST Flash error
Integrity check of Flash memory
POST HPS error
High pressure shutdown error
POST IB error
Infusion Bubble Detector error
POST MD error
Motor drive error
POST MOC error
Motor overcurrent shutdown error
POST PS error
Pressure Sensor error
POST PT error
Perfusion temperature sensor error
POST RAM error
Integrity check of Random Access Memory
POST ROM error
Integrity check of Read Only Memory
POST UB error
Upstream Bubble Detector error
POST WD error
Watchdog error