16 / 23
i button
CUSTOMER SERVICE
i
GEBRUEDER HEYL
i
i
E
E
i
ERROR REGISTRATION
st
i E
Error display, if it is programmed as
alarm
MODE OF OPERATION
Displayed unit
PROGRAM VALUES
st
i E
Measuring points 1 or 2
Flush times/interval
Limit values
INDICATOR TH LEVEL 030%
Residual analyses 1386
Reagent B level
030%
Reagent A level
030%
OPERATING VALUES
st
i
Res. operat. time 00126h
D-31135 HILDESHEIM
TEL. 05121-7609-0
Maintenance
Program values
Error registration
CUSTOMER SERVICE
Operating values
INFORMATION
st
i E
Maintenance interval 000d
MAINTENANCE
i
Maintenance in
000d
Up to 20
error messages
Low water pressure
Ff. optics
Ff. outlet to drain
Ff. dosing pump
Mf. dirtiness
Transfer error
power failure 24 V
REAGENT Cl LOW LEVEL
Indicator TH low level
Power failure
TH: 0.20 °dH
Cl: 0.24 mg/l
^1: 0.20 °dH
^2: 0.40 mg/l
E
i
1
2
3
4
5
6
i
E
POWER FAILURE
i
E
Reagent C level
030%
4 lines are displayed
(black scale)
Water quantity
0063m
3
Software version 48EXXX
Operating time
00629h
i
E
to
From
Information Menu "i"
Structure of the "i" - menu
The information menu can be used
to call up the active settings and
status of the unit, the error
registration, the date for the next
maintenance and the customer
service address.
Call
(1)
Call up the information menu "i" by
pressing the "i" key.
Customer Service
(2)
Display of the customer service
address or, e.g. of a service
telephone number.
These three lines can be
independently programmed in the
basic program (
protected by a
password
):
"M"
º
BASIC PROGRAM
º
CUSTOMER SERVICE
Operating values
(3)
Display of the current values.
Program values
(4)
Call up the menu item "program values" by pressing the arrow keys. By pressing "ENTER" the list with the set values
can be opened. The active setting of a parameter can be queried by pressing "ENTER".
The selected function is marked by a star.
(In this context there are no active lines.)
Error registration
(5)
Open the error registration by pressing the "i" key and then "ENTER". The error registration is a list of the errors or
status, that have been programmed as faults. This list is lost during a power failure and the recording is started afresh
Provided no faults have accumulated since the start-up, the display will show the last time the unit was switched on, e.g.:
POWER FAILURE
from
16.06.99 06:56
to
16.06.99 07:09
Maintenance
(6)
Display of the date of the next maintenance and of the programmed maintenance interval.
The maintenance interval can be set in the basic program (
protected by a password
):
"M"
º
BASIC PROGRAM
º
SERVICE II
For further maintenance information refer to page 21 onward.