17 / 23
Prozeß
M - key
SERVICE I
s
st
t
M E
INPUT REAGENT
Manual operation
Confirm maintenance
DATE / TIME OF DAY
s
st
t
M E
DATE 11.01.00
Time of day 23.59
PROGRAM MENU
Testomat 2000 TH/Cl
BASIC PROGRAM
INPUT REAGENTS
s
st
t
M E
INDICATOR TH
LEVEL: 100%
Reagent A
level: 100%
Reagent B
level: 100%
E
Day of week Monday
MANUAL OPERATION
s
st
t
M E
FLUSH
*
Drain chamber
Fill chamber
Flush chamber
TH: 0.20 °dH
Cl: 0.24 mg/l
^1: 0.20 °dH
^2: 0.40 mg/l
NOTE !
M E
Manual function without
monitoring!
go on with enter key
DIAGNOSIS
s
st
t
M E
OUTPUT LV1
Output LV2
Output MP 1/2
Output Flush
Output AUX
Output Alarm
*
Output Maintenance
*
Output OUT
255
Input Start
Input Stop
Diagnosis
E
1
2
3
4
4
5
6
7
8
10
11
12
E
MAINTENANCE
M E
Confirm the course with enter!
E
PROGRAM
s
st
t
M E
SERVICE I
Basic program
Date/Time of day
E
E
Input IN1
*
Input IN2
*
E
E
Description on the next
page
Reagent C
level: 100%
Program Menu "M"
Call
(1)
The program menu "M" is selected by
pressing the "M" key
Except for the basic programming, all
functions can be called up without password
protection.
Service I
(2)
Input reagent
(3)
The new filling level has to be entered after
every
refill or replacement of the reagent
bottle. As soon as the menu item for the
filling level "Reagent X filling (0 - 100%)" has
been selected by pressing "ENTER", the
value is preset to 100%. Confirm the
connection of a full bottle pressing "ENTER".
If only a partially filled bottle is connected,
enter the corresponding % value.
Manual operation
(4)
After the information message (4) has been
confirmed by pressing ENTER, the desired
function can be selected using the arrow
keys and then activated by pressing
"ENTER".
These functions are used for monitoring the
operation and for commissioning.
L
All manual functions can only be selected during an interval between two analyses. No analyses are carried out
during manual operation. All signal inputs and outputs are locked.
Flush
(5)
Start the flushing of the sampling tube through the internal valve by pressing "ENTER". This function is terminated by
pressing the "ENTER" key again.
Flush chamber
(6)
By pressing "ENTER", the measuring chamber is flushed once.
Drain chamber
(7)
The outlet valve to drain the water from the measuring chamber is opened by pressing "ENTER".
This function is terminated by pressing the "ENTER" key again.
Fill chamber
(8)
The measuring chamber is filled by pressing "ENTER".