19
2.7.5
Waste Heat/Required Cooling Performance
The TCS SP5 system has a maximum power consumption of 3.2 kW (VIS system) or 6.2 kW
(MP system).
The TCS SP5 X has a maximum power consumption of 3.4 kW.
For the specifications of external lasers such as UV and MP lasers, please refer to the
manufacturer's documentation.
2.8 Features
2.8.1
Overview of Usable VIS/UV Lasers
The Leica TCS SP5 features a combination of the lasers listed below.
Laser type
Wavelength
[nm]
Maximum
luminous power
at laser output
[mW]
Maximum
luminous power
in focal plane
[mW]
Pulse duration
Diode 405
405
< 120
< 7
Continuous wave
(cw)
Diode 405 p
405
< 5 (mean
power)
< 0.3 (mean
power)
pulsed, 60 ps
DPSS 445
445
< 75
< 7
Continuous wave
(cw)
Ar
458, 476, 488,
496, 514
< 200
< 50
Continuous wave
(cw)
Ar, external
458, 476, 488,
496, 514
< 500
< 125
Continuous wave
(cw)
HeNe
543
< 1.5
< 0.5
Continuous wave
(cw)
DPSS 561
561
< 100
< 12
Continuous wave
(cw)
HeNe
594
< 4
< 1
Continuous wave
(cw)
HeNe
633
< 15
< 5
Continuous wave
(cw)
UV, external
355
< 500
< 18
Continuous wave
(cw)
Table 1: Usable lasers for TCS SP5 (without MP)