18
2.7.3
Requirements Regarding Ambient Conditions
Do not expose the system to drafts.
Ensure that the system is not installed next to air conditioners or ventilation systems. For this
reason, the installation location should be carefully planned.
Ensure that the environment is as dust-free as possible.
Also read the notes on protection against dust in Chapter 11 Care and Maintenance
Installing the system in darkened rooms is also advisable.
The system requires doors with inside widths of 1.0 m for installation, maintenance and
transport.
With regard to the load-bearing capacity of the floor, note that the system will apply a static
load of 200 kg/m².
Ensure that the environment is as vibration-free as possible.
2.7.4
Permitted Ambient Conditions
Permissible temperature range for operation:
+18 to +25
º
C
Temperature range for optimum optical
behavior:
+22 °C ±1 °C
Permitted relative humidity:
20 - 80% (non-condensing)
Permitted vibrations:
Frequency range [5 Hz–30 Hz]:
Frequency range [> 30 Hz]:
< 30
μ
m/s (RMS)
< 60
μ
m/s (RMS)
Pollution degree:
Class 2