© Leica
TC605/TC805/TC905/L-1.1.1en
147
Error reports and warnings
Remedy
Clear internal memory
Enter a valid value
Change battery
Cool or warm the
instrument
Inspect cables and
interface parameters
Inspect cables, and
functionality of external
recording device. Inspect
baud rate
Repeat using a lower baud
rate
Redetermine value, or
contact service
department
Contact service
department
If fault re-occurs, contact
service department
Inspect pointing. Check
that distance is not too
great
Cause
Internal memory is full
Non-valid value was
entered
Battery exhausted
The inside of the
instrument is too hot or
too cold
Parity error at interface
The system is receiving
no response from the
interface
The data are being
transmitted too quickly
V-index measured, or
line-of-sight error,
is > 54' (> 1 gon)
Measuring error angle
scanning
System error in
compensator
EDM signal faint, absent
or distorted
Error report
Full
Invalid Value
Battery Empty
Temperature
Parity Error
RS232 Timeout
RS232 Overflow
Value >1 gon
Angle Error
System error
EDM Signal
02
03
12
19
21
22
24
44
50
51
55