Care and maintenance
32
Leica M220 F12 / Ref. 10 716 922 / Version 03
10.4 Changing the fuse
WARNING
Risk of death from electrical shock!
X
Disconnect the power cable from the instrument power
socket before changing fuses.
1
2
The fuses are located in a fuse holder (2) in the instrument power
socket.
X
Using a screwdriver, pry out the fuse holder (2), then pull it all
the way out by hand.
X
Change out the fuses (1).
Use only 6.3 AH time-lag fuses.
X
Replace the fuse holder (2) and press it in all the way using your
hand.
10.5 Replacing the LED module
CAUTION
Burn hazard!
The LED module get very hot.
X
Check that the LED module have cooled before you
remove it.
The Leica M220 F12 does not have a built-in auxiliary lamp.
A replacement LED module should always be kept at hand. If
an LED burns out, another spare LED module should be
purchased to be kept at hand by the user.
Use only Leica original replacement LED module.
The replacement of the LED module of the illumination is
described in detail in the replacement instructions that
accompany the LED replacement module (10 448 374 Leica
order number).
10.6 Checking the function of the LED
illumination
See chapter 7.13.1.