Instrument Setup
Leica M220 F12 / Ref. 10 716 922 / Version 03 51
5 Instrument Setup
5.1 Preparatory work
1
X
Make sure that all parts are seated firmly.
X
Connect the Leica M220 F12 surgical microscope to the power
supply and switch on the power switch (1).
X
Balance the swing arm (see chapter 7.10 in the user manual).
5.2 Checking the functions
Check the following functions in accordance with the user manual:
X
Release all brakes of the Leica M220 F12 and move the stand
over the entire range of movement.
Result: The brakes disengage; the microscope can be moved over
the entire range of movement freely and without noise.
X
Move down the X/Y ranges and focus ranges using the
footswitch.
Result: Even, quiet movement over the entire range.
X
Adjust the magnification across the entire range.
Result: Adjustment possible over the entire range, no unwanted
noise.
X
Switch on the illumination and test it across the entire
brightness range.
Result: The illumination functions properly and can be adjusted.
Otherwise, please contact Leica Microsystems service with regard
to this matter.