Preparation before surgery
Leica M220 F12 / Ref. 10 716 922 / Version 03 19
7.5 Adjusting second observer tube
1
X
Insert stereo attachment/second observer tube into beam
splitter.
X
Tighten the rotary ring (1).
Stereo attachment for second observer
The dual stereo attachment can be attached to the left or right on
the beam splitter and turned any direction.
Adjusting the second observer tubes
2
X
Rotate the monocular second observer tube in the desired
direction.
X
Adjust the diopter setting at the eyepiece.
X
Correct the image with the knurled ring (2).
Adjusting the stereo attachment for second observer
X
Turn the attachment for second observer in the desired
direction.
X
Align the binocular tube horizontally.
X
Adjust the diopter setting at the eyepiece.
X
Correct the image with the knurled ring (2).
7.6 Installing documentation
accessories
WARNING
Risk of injury from parts falling down!
X
Complete all preparations and adjustments on the
optics carrier before the operation.
X
Never rebalance or re-equip the optical components and
accessories with the instrument over the operation area.
X
Before re-equipping, always lock the swing arm.
X
Check that the optical components and accessories fit
well and are fastened securely before the operation.
X
Before changing equipment during an operation, swivel
the microscope away from the operation area.
1
2
1 Adjusting the magnification
2 Focusing knob
Fitting the phototube
X
Fasten the phototube to the documentation port of the 0°
assistant attachment or to the beam splitter.
X
Secure the camera, complete with adapter, in the phototube.
Tighten the clamping screw.
Please refer to section 9.2 for a list of video accessories.