What to do if ...?
Leica M220 F12 / Ref. 10 716 922 / Version 03 37
Malfunction
Possible cause
Remedy
Magnification changer does not work.
Magnification changer locked.
X
Adjust the magnification manually on the
emergency drive (see chapter 6.5).
Motor/footswitch defective.
X
Carry out diagnostics according to
chapter 10.7.
In the event of a defect, contact Leica
Microsystems service.
Focus is not working.
Motor/footswitch defective.
X
Carry out diagnostics according to
chapter 10.7.
In the event of a defect, contact Leica
Microsystems service.
XY movement is not working.
Motor/footswitch defective.
X
Carry out diagnostics according to
chapter 10.7.
In the event of a defect, contact Leica
Microsystems service.
12.3 TV, photography
Malfunction
Possible cause
Remedy
The image on the monitor is too dark.
The video camera and/or monitor is/are not set
correctly.
X
Optimize the settings for the camera and/or
monitor (see manufacturer's operating
instructions).
Filter/diaphragm is set incorrectly.
X
Adjust the brightness or change the filter/
diaphragm.
The photos are unsharp.
The parfocality of the microscope is not set
properly.
X
Check the parfocality of the microscope (see
chapter 7.9).
Specimen is out of focus.
The specimen is not precisely focused.
X
Focus accurately, insert graticule if necessary.
If your instrument has a malfunction that is not described
here, please contact your Leica representative.