Preparation before surgery
18
Leica M220 F12 / Ref. 10 716 922 / Version 03
Installing the protective objective glass
2
3
4
5
X
Fit the objective protection glass holder (2) to the objective and
position it so that the marking (3) faces back.
X
Hold the objective protection glass (5) in position and turn
slightly to the right.
The protective objective glass engages, the marks (3) and (4) are
positioned one above the other.
The protective objective glass is not autoclavable.
Inserting additional filters
The Leica M220 F12 features two slots for additional filters.
You can use temperature-resistant filters with a diameter of
32 mm. For further information, please contact your Leica
Microsystems representative.
1
X
Pull out the 2 empty filter holders (1).
X
Fasten the filter in the filter holder.
X
Reinsert the filter holder (1).
7.4 Installing the adapter for
accessories
Installing the beam splitter/stereo adapter
1
X
Unscrew the clamping screw
(1).
X
Push the beam splitter/stereo
adapter into the dovetail ring.
X
Tighten the clamping screw.
Beam splitter with 50/50 % observation, alternative: beam
splitter with 70/30 % observation
Installing the adapter
1
X
Insert the adapter in the beam splitter.
X
Tighten the rotary ring (1).