Preparation before surgery
Leica M220 F12 / Ref. 10 716 922 / Version 03 23
7.13 Checks before operation
7.13.1 Checking the function of the LED
illumination
WARNING
Failure of the illumination can be dangerous for the patient!
X
Keep a compatible replacement LED module ready.
1
X
Switch on the microscope using the power switch.
The LED is illuminated.
X
Turn the button (1) to the right.
This increases the brightness.
X
Test the illumination across the entire brightness range.
7.13.2 Checking the focus drive
See chapter 8.5.
7.13.3 Check the fastening of the power cable
X
Check if the power cable is fasten to the swing arm using cable
ties (see chapter 2.4.1 in the installation manual).