16
Instructions for Use V 1.7 Rev. J – 12/2017
4. Setting up the instrument
4.3 Unpacking and installation
(Continued)
Fig.5
5
The packaging must be retained for
the duration of the warranty period.
To return the instrument, follow the
instructions above in reverse order.
4. Ensure that when removing the instrument (
5
)
from the pallet this is carried out by two people
lifting four lower corners of the housing base
(
Fig. 5
).
5. Place the instrument on a stable laboratory
table.
6. Remove the accessories from the accessory
box (
6
) on the base of the pallet.
Fig.6
6
4.4 Moving the instrument
Do not move the instrument during operation.
Before moving the instrument, make sure that there is no specimen blocks on the cold
plate, the instrument is at an ambient temperature, and the power cord is disconnected
from the power supply.
Do not touch the metal parts of the compressor air outlet (1) on the rear panel.
It is mandatory to observe a waiting time of 4 hours before the instrument is switched
on.
Hold the instrument at the front and rear part of
the lower housing base and move.
1
Fig.7