HistoCore Arcadia C – Cold Plate
4. Setting up the instrument
•
Stable, vibration-free laboratory table with horizontal, flat table top, as far as possible vi-
bration-free ground.
•
No direct sunlight or strong temperature fluctuations. Room temperature consistently be-
tween +20 °C and +30 °C.
•
Relative air humidity maximum 80 %, non-condensing.
•
The instrument should be set up in such a way that the air circulation is not impaired.
•
The instrument must be installed in a place that ensures an easy disconnection from the
power supply. The power cable must be in a place that can be easily reached.
4.1 Site requirement
At a room temperature of > +30 °C, the working temperature of the cold plate of -6°C
may not be reached at all points.
To ensure proper function and an easy disconnection of the power cable from the in-
strument, there must be gap of at least 15 cm behind the instrument. Failure to observe
this distance may result in serious damage to the refrigeration unit of the device. The
instrument should not be operated in hazardous locations.
4.2 Standard delivery—packing list
Qty
Designation
Order No.
1
Basic unit HistoCore Arcadia C,
220-240 VAC
14 0393 53599
220-240 VAC, China
14 0393 54102
110-120 VAC
14 0393 54092
100 VAC
14 0393 54091
220-240 VAC
14 0393 57262
220-240 VAC, China
14 0393 57263
110-120 VAC
14 0393 57261
100 VAC
14 0393 57260
4
Sets of spare fuses: 220-240 VAC, 5A 250 V
14 6000 05015
100-120 VAC, 10A 250V
14 6000 05078
1
Instructions for Use (printed German/English with language CD 14 0393 80200)
14 0393 80001