11
HistoCore Arcadia C – Cold Plate
3. Instrument Components and Specifications
3.3 Technical Data
General data
Voltage of power supply
100 VAC, 110-120 VAC, 220-240 VAC, 50/60Hz
Fuse
Time-lag fuses 5 x 20 mm
•
220-240 VAC: 2xT5A, 250V
•
100-120 VAC: 2xT10A, 250V
Nominal current
5 A max.
Maximum start-up current (5 s)
25 A
Environmental operating temperature range
+20 °C to +30 °C
Operating temperatures
-6 °C
Environmental relative humidity
20 to 80 % - non-condensing
Environmental operating altitude
Up to 2000 m
Permissible temperature range
during storage
+5 °C to +50 °C
Permissible temperature range
during transport
-29 °C to +50 °C
Permissible humidity range
during storage and transport
10 to 85 % - non-condensing
Electromagnetic environment
Basic electromagnetic environment
IEC 61010 classification
Protection class 1
Pollution degree
2
IP protection class (IEC 60529)
IP20
Refrigeration unit
Refrigeration capacity*
158 W( at 50Hz) ;185 W( at 60Hz)
Safety factor
3
Refrigerant
R 134a
Compressor oil
150 +10/-5 ml Ester RL7H, ISO 7
* according to ASHRAE, condensing temperature: 54.4 °C, evaporating temperature: -23.3 °C
Coolant
115 g ± 2 g