3.
Face I measurement
Aim the telescope accurately at
a target at about 100 m dis-
tance (or at least 20 m). The
target must be positioned at
least 27°/30 gon above or
beneath the horizontal plane.
The procedure can be started in
any telescope face.
0016435_001
V=90°
+ 27°
- 27°
4.
Measure
to measure and to
continue to the next screen.
Motorised instruments change
automatically to the other face.
Non-motorised instruments
guide to the other face.
☞
The fine pointing
must be performed
manually in both
faces.
0016436_001
180°
180°
5.
Face II measurement
Measure
to measure the same target in the other face and to calcu-
late the tilting axis error.
☞
If the error is bigger than the predefined limit, the procedure must
be repeated. The tilting axis measurements of the current run are
then rejected and not averaged with the results from previous runs.
6.
Adjustment Status
Number of measurements
: Shows the number of runs completed.
One run consists of a measurement in face I and face II.
a T-axis quality (1
σ
)
: shows the standard deviation of the deter-
mined tilting axis error. The standard deviation can be calculated
from the second run onwards.
☞
Measure at least two runs.
7.
Next
to continue with the check & adjust procedure.
8.
Select
Add another calibration loop
if more runs have to be
added.
Next
and continue with step 3.
OR
Select
Finish the calibration & store the results
to finish the cali-
bration process. No more runs can be added later.
Next
to view the
adjustment results.
9.
Select
Finish
to accept the results. No more runs can be added later.