5.6
Adjusting the Circular Level of the Prism Pole
1.
Suspend a plumb line.
1
2
4b
4a
TS_080
2.
Use a pole bipod, to align the
prism pole parallel to the plumb
line.
3.
Check the position of the circular
level on the prism pole.
4.
a
If the circular level is cen-
tred, no adjustment is nec-
essary.
b
If the circular level is not
centred, use an allen key
to centre it with the
adjustment screws.
☞
After the adjustments, all adjusting screws must have the same
tightening tension and no adjusting screw should be loose.
5.7
Inspecting the Laser Plummet of the Instrument
The laser plummet is located in the vertical axis of the instrument. Under nor-
mal conditions of use, the laser plummet does not need adjusting. If an adjust-
ment is necessary due to external influences, return the instrument to any
Leica Geosystems authorised service workshop.
0016438_001
3
4
5
360°
3 mm/1.5 m
Ø 2.5 mm / 1.5 m
1
The following table explains the most common settings.
1.
Set up the instrument on the tripod approximately 1.5 m above the
ground and level up.
2.
Select
Settings
\
TS instrument
\
Level & compensator
to access
the
Level & Compensator
panel.
The laser plummet is switched on when the
Level & Compensator
panel is entered. Adjust the laser plummet intensity.
☞
Inspection of the laser plummet should be carried out on a
bright, smooth and horizontal surface, such as a sheet of
paper.
3.
Mark the centre of the red laser dot on the ground.
4.
Turn the instrument slowly through 360°, carefully observing the
movement of the red laser dot.
Adjusting the circular
level step-by-step
☞
Inspecting the laser
plummet step-by-step
44
Check & Adjust