Select the option:
Check & adjust the compensator, index error
& line of sight error
3.
Next
4.
Face I measurement
☞
Use a clean Leica standard prism as the target. Do not use
a 360° prism.
5.
Aim the telescope accurately at a
target at about 100 m distance.
The target must be positioned
within ±9°/±10 gon of the hori-
zontal plane.
The procedure can be started in
any face.
0016433_001
± 9°
~ 100 m
6.
Measure
to measure and to con-
tinue to the next screen.
Non-motorised instruments guide
to the other face.
☞
The fine pointing has
to be performed man-
ually in both faces.
0016434_001
180°
180°
7.
Face II measurement
Measure
to measure the same target in the other face and to calcu-
late the instrument errors.
☞
If one or more errors are bigger than the predefined limits, the pro-
cedure must be repeated. All measurements of the current run are
rejected and none of them is averaged with the results from previous
runs.
8.
Adjustment Status
Number of measurements
: Shows the number of runs completed.
One run consists of a measurement in face I and face II.
l Component quality (1
σ
)
: and similar lines show the standard
deviations of the determined adjustment errors. The standard devia-
tions can be calculated from the second run onwards.
☞
Measure at least two runs.
9.
Next
to continue with the check & adjust procedure.
40
Check & Adjust