Page 51 Operating Manual EM CPD300 12/18
17. Warnings and Troubleshooting
The list below summarizes Warnings (W...).
Regarding most of the displayed information, just follow the instructions; in the case
of error messages, contacting your local Leica Service is required.
Code
Warning
Cause
Action
10E11
Com connection
default
Call Leica service
10E12
Lid open
Process can not start if lid
open
Close lid
10E13
Separator missing
Process can not start or
continue if separator is
missing
Insert separator
10E14
Separator full
Process can not start or
continue if separator is full
Empty separator
10E16
Heatup error
Pressure too low
Check CO
2
-bottle
10E17
Timeout CO
2
-IN
CO
2
-bottle empty or poral
inlet filter is clogged
Check CO
2
-bottle
or exchange poral
inlet filter
100E3
Pressure sensor
failure
Call Leica service
100E4
Pressure fault
Poral outlet filter blocked
Exchange poral
outlet filter
100E5
Temperature sensor
failure
Call Leica service