Page 42 Operating Manual EM CPD300 12/18
13. Adjustments of Pressure Threshold for Bottle Empty
Function
The bottle empty function was developed to protect the samples if the CO
2
bottle
becomes empty during a run. When the warning occurs, all valves will be closed so
that the pressure chamber is sealed, and the empty bottle can be exchanged with
reduced possibility of sample damage. The threshold for this function has to be
adapted to the CO
2
temperature. See list below. Green marked values indicate
optimal working temperature and pressure range.
CO
2
-Temperature (°C)
Recommended
threshold for pressure
(bar)
Pressure of full CO
2
-
Bottle (bar)
14
47
50
15
48
51
16
49
52
18
52
55
20
54
57
22
57
60
24
60
63
25
61
64
26
63
66
28
66
69