Page 37 Operating Manual EM CPD300 12/18
8.2
Filling the chamber with exchange fluid
Do not forget adjustment of the pressure threshold for the bottle empty function (See
page 38-39)!
Open the screw-on cover of the sample chamber by turning counter clockwise.
Before filling the sample chamber with exchange fluid, place the magnetic stirrer rod
together with the spacer into bottom of the sample chamber.
Fill exchange fluid into the sample chamber until the stirrer and spacer is covered.
Fill the sample chamber with exchange fluid to such a level that the specimen holder
with specimen will be completely submerged when inserted.
8.3
Inserting of Specimen into Sample Chamber
Open lid and transfer prepared specimen into a suitable sample holder.
Quickly insert the sample holder with the specimen into the sample chamber filled
with exchange fluid. Check if sample holder is still covered with exchange fluid.
8.4
Cooling of the Sample Chamber
After tightly closing the screw-on cover of the sample pressure chamber, press Start.
Press the cooling button and wait until pressure chamber cools up to the preset
cooling temperature (15°C factory setting). The cooling temperature depends on the
bottle temperature. Please see sub chapter “7.6.1 CO2-Bottle Temperature /
Pressure Function” to select the appropriate cooling temperature. The cooling unit
automatically keeps the temperature at the preset cooling temperature or below.
Spacer with magnetic Stirrer
Summary of Contents for EM CPD300
Page 1: ...From Eye to Insight OPERATING MANUAL EM CPD300 167180032 Version 12 18 ...
Page 30: ...Page 27 Operating Manual EM CPD300 12 18 Press Start to activate program Test ...
Page 48: ...Page 45 Operating Manual EM CPD300 12 18 After software update press back ...
Page 52: ...Page 49 Operating Manual EM CPD300 12 18 Remove poral filter with spring from the nut ...