Installation
3
Note
• Carefully check the delivery against the packing list, delivery note and your order.
• Should there be any discrepancy, please contact the Leica sales unit handling your order or your
Leica dealer.
3.3
Setting up the instrument/Site requirements
Warning
• Do not operate the instrument in rooms with explosion hazard.
• To ensure proper function of the instrument, it must be set up while maintaining a minimum
distance of 10 cm from walls and furniture.
• The instrument requires an installation area of approx. 150 x 200 mm.
• The substrate must have a sufficient load capacity and rigidity with respect to the weight of the
instrument.
• Relative humidity: 20 to 80 % – non-condensing.
• Room temperature consistently b18 °C and +40 °C.
• Elevation: up to a max. 2000 m above sea level.
• The instrument is designed for indoor use only.
• Avoid vibrations, direct sunlight and heavy variation in temperature!
• The power supply must be within the range of the length of the power cable. No extension cable may
be connected.
• The instrument
MUST
be connected to a grounded power socket.
• Use only one of the provided power cables that is intended for the local power supply.
3.4
Electrical connection
The Leica EG F can be connected to different electrical power systems (voltage and frequency-
dependent, refer to Chapter
Please observe the following notes to prevent damage to the instrument:
Warning
• The instrument
MUST
be connected to a grounded mains socket.
• Only the power cord intended for the local power supply (socket) may be used.
• Do not use an extension cord!