53
Measuring progress
Different screens are displayed while the measure-
remnts are in progress, depending on the measure-
ment mode selected.
Single measurement
Measurement time is very short. An hour glass is
displayed to indicate that measurement is in pro-
gress.
Multiple measurements
DNA-Dde 8
All the important information required to evaluate
multiple measurements is placed in a single display.
Count
Amount of measurements made (n).
Staff
Current staff height according to mode (mean,
median or single measurement) after "n" measure-
ments.
sDev
Current standard deviation of a single measurement
after "n" measurements.
sDevM
Current standard deviation of the mean after "n"
measurements.
Spread
Spread of the single measurements after "n" mea-
surements.
Spread =
largest measured value - smallest measured value
After the last measurement the display
remains static for approx. 3 seconds.
Shortens the display time.
Manual cancellation of multiple measurements
The last valid measured value is accepted
and stored.
Measuring...
Mode : Median
Count: 5
Staff: 2.8005 m
sDev : 0.0003 m
sDevM: 0.0001 m
Spread: 0.0007 m
EC
or