47
Leica CM1510 S – Cryostat
Before replacing the lamp, turn off
instrument mains switch and pull
mains plug!
•
Hold sliding window (
1
) at grip (
2
), lift slightly
and pull it out to the front.
If the lamp is broken, it must be re-
placed by the technical service -
high risk of injury!
Removing the lamp
•
Lamp (
3
) is mounted behind glare shield (
4
)
and therefore not visible.
•
Touch the lamp for better orientation.
•
Lightly tilt the fluorescent tube down to the
left and pull it out of clip (
5
).
•
Hold lamp with both hands and pull it to the
left, out of socket (
6
).
Do not use any lamp type other than
the one inserted by the manufac-
turer!
8.
Cleaning and maintenance
5
4
3
6
Fig. 33
Fig. 34
Installing the new lamp
Lamp type:
(230 V/ 50 Hz) OSRAM DULUX S - 11 W
(115 V/60 Hz) OSRAM DULUX S - 13 W
•
Hold lamp in correct mounting position -
see Fig. 34 - and push to the right until it en-
gages in the socket.
•
Lightly push the lamp upward to engage in
the clip.
•
Reinsert sliding window.
•
Reconnect instrument to mains power and
turn it back on.
8.8
Replacing the lamp
1
2
Fig. 32