Instrument components and specifications
Microtome
Rotary microtome
Section thickness setting
1 -60 µm;
in 1-µm steps from 1 to 10 µm
in 2-µm steps from 10 to 20 µm
in 5-µm steps from 20 to 60 µm
Total horizontal specimen feed
25 mm
Vertical stroke
59 mm
Maximum specimen size
55 x 55 mm
Specimen orientation
8° (x-, y-, z-axis)
Electrical coarse feed
slow
0.2
µm/s
fast
0.7
µm/s
Lamp
50 Hz version:
Osram Dulux S 11 W/21, light color LUMILUX bright white
60 Hz version:
Osram Dulux S 13 W/21, light color LUMILUX bright white
Cryocabinet
Dimensions
Width (w/o handwheel)
880 mm
Width (with handwheel)
680 mm
Depth
680 mm
Height (arm rests)
1010 mm
Height (overall)
1175 mm
Weight
(incl. microtome)
approx. 135 kg
Please refer to chapter 4.1 - ‘Installa-
tion site requirements’!
All CM 1510 S instrument types require the following secondary fuses:
F4: T0.5 A 6.3 x 32 mm;
Fa. Schurter, Typ FST;
Fa. Littlefuse, Typ 313
F5: T1.0 A 6.3 x 32 mm;
Fa. Schurter, Typ FST;
Fa. Littlefuse, Typ 313
F6: T2.0 A 6.3 x 32 mm;
Fa. Schurter, Typ FST;
Fa. Littlefuse, Typ 313
F7: T4.0 A 6.3 x 32 mm;
Fa. Schurter, Typ FST;
Fa. Littlefuse, Typ 313
F8: T2.0 A 6.3 x 32 mm;
Fa. Schurter, Typ FST;
Fa. Littlefuse, Typ 313