Installation site requirements
Do not operate in rooms with explosion hazard!
To ensure the instrument works trouble-free, the instrument must be set up with at least 10
cm distance from all walls and furniture.
Do not place anything next to the compressor ventilation grid (right side of the cabinet) to
ensure adequate ventilation!
The place of installation must meet the following requirements:
•
no direct sunlight,
•
mains power socket at a distance no greater than 2.5 m (length of
power cord = 2.5 m) - no extension cords must be used!
•
no drafts (air condition outlets etc.),
•
even floor,
•
virtually vibration-free floor,
•
handwheel freely and easily accessible,
•
room temperature maximum 35 °C, (< 22 °C recommended),
•
relative humidity max. 60 %,
•
only for indoor use.
High room temperature and excessive air humidity at the installation site affect the cooling
capacity of the cryostat.