12
INSTALLATION (Part 5 of 5)
)
and stabilize for at least 90 minutes. If the thermocouple reading and the controller actual temp differ by more than a
couple degrees, the control must be calibrated. To begin, make a note of the temp difference between the thermocouple
and controller actual temp, then turn control power button off and simultaneously press-and-hold the set temp up/down
arrows for 3 seconds until a value appears in set point display. Use set point up or down arrow to add or subtract difference
in temp. (EXAMPLE: If thermocouple is 10
o
higher than actual temp, add 10. If thermocouple is 10
o
lower than actual temp,
subtract 10.) Turn control button on and re-check thermocouple against controller actual temp. Repeat this procedure
until both temps match.
Final Inspection
Verify the following:
1.
All screws are tight
2.
All electrical connections are tight
3.
All plumbing connections are tight and do not leak
4.
Fingerprints, dirt and sticker residue have been cleaned from oven