11
Technical data
11.1
General data
The device sound pressure level is below 70 dB. According to EC
Directive 2006/42/EC the sound pressure level of the devices is
therefore not specified further.
Specification
Value Unit
Installation
Interior rooms
Installation altitude above sea level
up to 2,000 m
Air humidity
maximum relative humidity of 80 % at 31 °C and
decreasing linearly to 50 % up to 40 °C
Ambient temperature range
5 – 40 °C
IP protection level
IP 32
Pollution degree
2
Clearance (front and back)
40 cm
Surge
Surge category II and transient surges according to
category II
Protection class for electrical equipment DIN EN
61 140 (VDE 0140-1)
1
Classification according to DIN 12 876-1 (class
designation/marking)
I/NFL
Display
7-segment, LED
Display resolution
0.1 °C
Setting resolution
0.1 °C
Temperature stability
±
0.5 K
Storage temperature range
5 – 40 °C
Transportation temperature range
-20 – 60 °C
Working temperature range
Dimensions (W x D x H)
Weight
°C
mm x mm x mm
kg
MC 250
-10 – 40
200 x 350 x 465
28
MC 350
-10 – 40
240 x 400 x 500
36
MC 600
-10 – 40
350 x 480 x 595
52
MC 1200
-10 – 40
450 x 550 x 650
64
V07
Microcool
50 / 57