2.
Select the following option to output an electrical signal for alarms and
errors.
3.
Select the following option to output an additional electrical signal for
warnings.
Wait approx. 4 seconds if you do not wish to accept the
specified value. The display will automatically return to the
basic display.
4.
Press the input button to confirm.
The value must be confirmed within 4 seconds of the last
entry. Otherwise, the display will return to the basic display.
6.8.2 Interface potential-free contact
n
The contacts may be loaded with a maximum voltage of 30 V direct
current (DC) and a maximum current of 0.2 A.
Fig. 23: Flange plug (front) in fault condition
View of the flange plug (front) or into the coupling socket on the
soldering side.
Good condition
n
Pin 1 and 2 are closed.
n
During trouble-free operation, the alarm output is in good state.
Fault condition
n
Pin 2 and 3 are closed.
n
The alarm output is in fault condition:
If the device is switched off,
after switching on, if a fault (e.g. low level) is already present,
during operation, if a fault occurs, and
for each event configured in the
Alarm Output
menu.
Fig. 21: Error and alarm option
Fig. 22: Option with additional warning
V07
Microcool
41 / 57